SOIL-DAQ™
CALIBRATION
Operator’s Manual EZ-DAQ™ version 1.4 23 of 25
Manually Adjusting Calibration Data.
The system allows manual input of A/D readings for re-entering calibration data or for
tweaking calibration data points to achieve proper calibration results.
2.0 Important Information
The above calibration steps have been provided for a yearly calibration check on the
DGSI EZ-DAQ touch screen.
Sensor types have been downloaded to the system by the factory before it shipped.
For example, sensor 1 is a 4-20 mA signal. This info has been coded into the touch
screen at the factory. A 200 mV/v sensor can’t be interchanged into this slot.
If a different sensor type is required, then factory recalibration is required.
Press here
for keypad.
Enter new desired A/D
number from keypad.
Press here